| MECHANISM | TECHICAL DESCRIPTION |
01 | Channel | Dual, isolated between channels |
02 | Dimensions | 107x62x27mm |
03 | Weight | Approximately 150 grams, battery included |
04 | Amplitude | Adjustable 0-60mA peak into 500 ohm load each channel,constant current |
05 | Carrier Frequency | 4000Hz, Fixed (CH1) |
06 | Modulating Frequency | 4-160bps, Adjustable (CH2) |
07 | IF Frequency Modes | 9 modes, 1 constant mode, 5 frequency shift modes, 3 auto sweep Modes |
08 | Constant Mode | 4-160bps, Adjustable |
09 | Auto Sweep | 4-45 bps 80-145 bps 4-SET bps
|
10 | Frequency Shift | Five modes, 1/1 abrupt shift 6/6 abrupt shift 6/6 ramped 10/10 abrupt shift 10/10 ramped bps adjustable |
11 | Sweep Time | 15 seconds |
12 | Frequency Shift Percent | Frequency shifts from 30% below the set frequency to 60% above and then return to 30% below the set frequency. |
13 | Output Configuration | Quad polar (4 electrodes) |
14 | Wave form | Symmetrical Balanced Sine Wave |
15 | Interference Pulse Frequency | 4-160 bps, Adjustable, 4 bps/step |
16 | Pulse Duration | 125 μs maximum |
17 | Patient Compliance Meter | Display total treatment time |
18 | Lock | Prevents the patient from accidentally changing any of the set parameters |
19 | Timer | 15, 30 minute selectable |
20 | LCD | Displays mode, bps rate, abrupt/ramp, timer, and channels |
21 | Max charge per Pulse | 15 Micro-coulombs Maximum |
22 | Low Battery | A low battery indicator will show when the battery is low. |
23 | Power Source | 9 Volt DC adapter, 500mA or 9 volt alkaline battery (not recommend due its short life) |
24 | Tolerance | There may be +/-10% tolerance of settings and 20% tolerance of output. |