Tens MED Tens Machine tens machines tens hire
Pay for your tens machine via paypalTENS Machine+ TENS Hire
Tens Machines EMS Units & Electrodeswww.TensMachines.co.uk
 
0 item(s) in cart / Total: £0.00
Home > EMG & ETS Stims
NeuroTrac™ MyoPlus2 - DUAL CHANNEL EMG + STIMULATION + ETS
Part Number NeuroTrac™ MyoPlus2
NeuroTrac™ MyoPlus2 - DUAL CHANNEL EMG + STIMULATION + ETS
NeuroTrac™ MyoPlus2 - Dual channel EMG (Biofeedback) With ETS
E-mail to friend    Review this product
Product Pricing
Our Price:  
£349.94
Product Availability:
IN STOCK + FREE UK Delivery
Enter Details Below
Add Extras 1 x Pack 50x50 Electrodes [+£4.94]
3 x Packs 50x50 Electrodes [+£12.50]
6 x Packs 50x50 Electrodes [+£22.50]
Pre-Tens Skin Preparation [+£4.94]
Veriprobe Womens Continence Probe [+£16.94]
Add Computer Software Neurotrac Software [+£240.00]
Enter Qty
 Description
Save 20% VAT Free if EligibleNeuroTrac™ MyoPlus2 DUAL CHANNEL EMG + STIMULATION + ETS

Dual channel EMG (Biofeedback) combined with NMS-STIM (Neuromuscular Stimulation) & ETS (EMG Triggered Stimulation).

Passive and Active phases: EMG and NMS in one programme

Manual and Automatic threshold control.

EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).

Stand alone device or used in conjunction with established multilingual software.
 
Features:
  • Dual channel EMG (Biofeedback) combined with NMS-STIM (Neuromuscular Stimulation) & ETS (EMG Triggered Stimulation).
  • Passive and Active phases: EMG and NMS in one programme
  • Manual and Automatic threshold control.
  • EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
  • Stand alone device or used in conjunction with established multilingual software.
  • Pre-set neuromuscular Stimulation modes for: Continence, Sports, Rehab.
  • Custom programmes enabling up to 5 phases (any of: STIM W/R-MODCONT- ALT, EMG, ETS) to be set up in each programme.
  • Multilingual LCD screen and voice prompts for Biofeedback.
  • For home users: Produces and prints (with the optional PC Software) the comprehensive patient progress reports on a day to day basis providing average readings (2 channels) on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
  • Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscularstimulation.
  • Multilingual LCD screen prompts.
  • 4 x AA batteries {No mains required}.
Programmes:
UROLOGY: p01-p19SPORTS: p01-p21REHABILITATION: p01-p08
P1 Pelvic floor pain
P2 Urge incontinence 1
P3 Stress incontinence 1
P4 Stress incontinence 2
P5 Frequency / Urge 1
P6 Frequency / Urge 2
P7 Frequency / Urge 3
P8 Lack of sensitivity
P9 Pelvic floor work out
P10 Building up endurance
P11 Relaxing the pelvic muscle
P12 Prolapse
P13 Stress incontinence 3
P14 Stress incontinence 4
P15 Urge incontinence 2
P16Urge incontinence 3
P17 Frequency
P18 Flaccid muscle
P19 Weak pelvic muscle
P1 Warm Up
P2 Active recovery
P3 Endurance
P4 Resistance 1
P5 Resistance 2
P6 Resistance 3
P7 Resistance strength 1
P8 Resistance strength 2
P9 Resistance strength 3
P10 Strength 1
P11 Strength 2
P12 Strength 3
P13 Max strength 1
P14 Max strength 2
P15 Max strength 3
P16 Max strength 4
P17 Explosive force 1
P18 Explosive force 2
P19 Hypertrophy
P20 Agonistic 1
P21 Agonistic 2
P1 Muscle growth
P2 Spinal column
P3 Parectic extremities
P4 Facial palsy 1
P5 Facial palsy 2
P6 Disuse atrophy
P7 Stroke 1
P8 Stroke 2
P9 Multiple sclerosis

CUSTOM PROGRAMME
STIM
EMG
 
Specifications:

EMG  

1.0 Dual channel EMG

  1. 1.1 EMG Range: 0.2 to 2000 ìV RMS (continuous)
  2. 1.2 Sensitivity: 0.l ìV RMS
  3. 1.3 Accuracy: 4% of ìV reading +/-0.3 ìV at 200 Hz
  4. 1.4 Selectable Bandpass filter - 3db Bandwidth,
    a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% - Reading
    below 235 microvolts
    10 Hz +/-3 Hz to 370 Hz +/- 10% - Reading
    above 235 microvolts
    b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
  5. 1.5 Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0. 1% accuracy)
  6. 1.6 Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
  7. 1.7 a. Battery: 1.5V, AA battery
    b. Optional Charger: CE approved.
  8. 1.8 Work / Rest periods: 2-99 seconds
  9. 1.9 Number of Trials: 1-99


2. STIM (Neuromuscular Stimulation)

  1. 2.1 2 channel Stimulator
  2. 2.2 Amplitude: 0-90 mA into 500 Ohm load - actual mA will tend to be lessthan indicated due to Electrode impedance: at 1000 Ohms load (Electrodesin poor condition) the maximum will be limited to 75 mA, at 1500 Ohms load the maximum will be limited to 50 mA.
  3. 2.3 Type: Constant current, maximum output
    voltage 70 Volts +5 / -10 Volts
  4. 2.4 Waveform: Symmetrical, rectangular,
    bi-phasic with net zero DC current
  5. 2.5 Pulse width selection: 50-450 ìS (2% accuracy)
  6. 2.6 Pulse rate selection: 2-100 Hz (2% accuracy)
  7. 2.7 Work / Rest periods: 2-99 seconds
  8. 2.8 Time 1 - 99 minutes
  9. 2.9 Ramp up time: 0.1 - 9.9 seconds
  10. 2.10 Preset and user programmable treatment Programs
  11. 2.11 Automatic output shut off with detection of open electrode above 0.5 mA
     
Battery: Low battery indication at 5.4 volts +/- 0.2 volts, automatic shut off when voltage dropps below the low indication. Replace the batteries immediately!
The device goes off automatically when not in use (energy saving): for example when in settings and no key pressed over 1 minute, when in stimulation mode (home screen) and all channels are 0mA.

Environmental Conditions for use:
+10 to +30 degrees Centigrade. 0-90% Humidity.
Environmental conditions for storage & transport:
-10 to +50 degrees Centigrade. 0-90% Humidity.
Physical:
Dimensions:
 Length 150 mm, Width 89 mm, Depth 35 mm.
Weight:
MyoPlus2 device: 0.16 Kg (without battaeries).
 
Software packNeurotrac Software
Neurotrac Software v4.00 or higher for Windows XP, Vista, Win7

The software that is needed to get the best performance from the Neurotrac ETS machine, Neurotrac Simplex and Neurotrac MyoPlus, Neurotrac MyoPlus2, Neurotrac MyoPlus4.

The Neurotrac Software has the following capabilities:

1.: Patients Database

2.: EMG Work/Rest Assessment

3.: EMG Games (applicable with Neurotrac MyoPlus4)

4.: Patient Clinical Report

5.: Patient Home Compliance Report

6.: MYO Controlled for multiphase visualisation of any MyoPlus programme, including ETS visualisation

7.: Neurotrac Controlled for ETS visualisation from ETS device

Minimum System Requirements:

- Pentium 150, 32 MB memory 50 MB of free Hard Drive space.

- 1 free USB port for USB-Fibre Optic adapter.

- PC Monitor 800 x 600 px or wider

- Windows / 2000 / XP / Vista operating systems

Accessories provided in the software pack:

1 Neurotrac Software Installation

1 Fibre Optic USB Adapter

1 Fibre Optic cable - connects Neurotrac devices to the Fibre Optic adapter

1 Manual

Product Accessories
Sony AA Alkaline Tens/Ems Batteries 4pk
Sony AA Alkaline Tens/Ems Batteries 4pk
Our Price: £3.50
 0 Review(s)
Uni-Stim Electrodes - Size of Pads: 30mm x 30mm Round
Uni-Stim Electrodes - Size of Pads: 30mm x 30mm Round
Our Price: £4.60
 0 Review(s)
Uni-Stim Electrodes - Size of Pads: 40mm x 40mm Square
Uni-Stim Electrodes - Size of Pads: 40mm x 40mm Square
Our Price: £4.85
 0 Review(s)
Uni-Stim Electrodes - Size of Pads: 50mm x 50mm Round
Uni-Stim Electrodes - Size of Pads: 50mm x 50mm Round
Our Price: £4.94
 0 Review(s)
Uni-Stim Electrodes - Size of Pads: 50mm x 50mm Square
Uni-Stim Electrodes - Size of Pads: 50mm x 50mm Square
Our Price: £4.94
 10 Review(s)
Veriprobe For Incontinence Tens Machine
Veriprobe For Incontinence Tens Machine
Our Price: £16.94
 0 Review(s)
  
Browse Similar Items
Incontinence Tens
Sports Tens / EMS